
resistance. The moving coil is connected between the load
terminals in such a way that it is passed through by a current
proportional to the load voltage.
The instantaneous torque that generates the angular
deflection of the mobile system can be found from the energy
of the field formed by both coils:
𝜏
= 𝑖
∙ 𝑖
∙
, (1)
Where 𝑖
is the current flowing through the fixed coil and
𝑖
through the moving one, 𝑑𝑀 𝑑𝜃
⁄
is the derivative of the
mutual inductance M with respect to the deflection angle 𝜃.
Considering that the time constant of the system is much
greater than the period of the electrical variables and that the
measurement reaches equilibrium when the mean deflecting
torque becomes equal to the restoring torque 𝑇
= 𝑘 ∙ 𝜃
where k is the spring constant, one obtains:
𝜃 =
∙
∫
𝑖
∙ 𝑖
∙ 𝑑𝑡 , (2)
Assuming, ideally, that the moving coil is purely resistive,
and substituting the current 𝑖
by 𝑢
𝑅
⁄
in (2) it is
concluded that the deflection is proportional to the power of
the circuit in which it is measured.
For the sinusoidal case with 𝑖
(
)
=
√
2 ∙ I
sin
(
𝜔𝑡 − 𝜑
)
and 𝑖
()
=
√
2 ∙ U
𝑅
sin
(
𝜔𝑡
)
substituting in (2):
𝜃 = 𝑘
∙ I
∙ U
∙ 𝑐𝑜𝑠𝜑 = 𝑘
∙ 𝑃 , (3)
Which allows to see that the deflection is proportional to
the active power.
The sources of error in electrodynamometer wattmeters
have been extensively studied. In [5] they are broken down
between errors due to voltage or current separately, those that
affect the product of both and those that affect the angle.
Among the former are magnetic impurities in the mobile
system that can cause deflections with currents only in the
fixed coil, induced or capacitive currents in the moving
system that increase with the square of the magnitude and
frequency of the current through the fixed coil. The errors due
to the U.I product, called errors at unity power factor,
generally are discarded as being negligible at frequencies up
to 3 kHz. In the case of phase angle dependent errors, the
causes include eddy currents in the fixed coils or skin effect.
Also, errors may be caused by capacitive couplings in the
fixed coils and by phase errors in the voltage circuit due to a
phase difference between the current that flows through the
mobile coil and the voltage applied to it.
Similarly, in [6], the analysis of audio frequency
measurements with electrodynamometer wattmeters is
carried out, emphasizing the shielding needs of the fixed and
mobile system to prevent external magnetic fields from
exerting a torque on the mobile system. The benefits of
shielding with a compensation capacitor and the use of
double shielding are evaluated.
In [7] the design and construction of an ED wattmeter for
accurate measurements in the range of direct current to 20
kHz is discussed. The main effort in the construction of this
instrument was devoted to obtain a sufficient torque with low
number of turns in the current and potential windings, paying
attention to their shapes and physical arrangement to
minimize the interaction effects of the moving and fixed coil
fields. To reduce capacitive influences at high frequencies,
the potential coils were wound with paper inserts. The current
coils are made with individually twisted and lacquered
conductors to avoid eddy currents.
Another design can be found in [8]. Where the moving
system of the comparator is in the form of a transverse
horizontal beam, acting as a balance, on the edges of which
plane coils of two identical electrodynamic transducers are
fastened, while their fixed coils are connected rigidly to a
beam parallel to the base of the comparator, the suspension
of the mobile system takes the form of four tension wires, the
electrodynamic transducer is in the form of flat coils. This
comparator is an extremely sensitive measuring instrument.
This enable only a small number of windings to be used in
the transducers and consequently it will have a low
inductance, giving the instrument a wide range of
measurement values and frequencies.
This brief review of history accounts for the effort to
optimize the system, make compensations and corrections in
order to increase the frequency range of electrodynamometer
type wattmeters. Without these improvements, the
instruments only guarantee the class of accuracy within the
kHz range. In Fig. 1 the percentage change in the reading due
to the frequency, presented by a manufacturer, can be
observed.
Fig. 1. Error increase with increasing frequency [Source: Yokogawa
Portable Wattmeters 2041, 2042].
It can be noted that a class 0.5 instrument, for example,
will measure at frequencies of the order of 2 kHz but with an
accuracy of the order of 1% of the full scale value.
III. THE ED WATTMETER WITH POLYHARMONICS
When it is required to measure power in systems with
distorted voltages and/or currents [9], power can be
expressed as:
𝑃 = 𝑈
∙ 𝐼
+
∑
𝑈
∙ 𝐼
∙ cos
𝜑
− 𝜑
, (4)
Where 𝑈
and 𝐼
are the DC components of the voltage
and current respectively, 𝑈
is the rms value of the k-order
harmonic component of the voltage, 𝐼
is the rms value of
the k-order current component, and 𝜑
, 𝜑
are the phases of
the components of order k of the voltage and current
respectively. The cross products between voltage and current
harmonics of different frequencies are zero due to
orthogonality between sinusoidal functions that are multiples
of the same frequency.
It can be noted that the total power within a given
bandwidth is the sum of the DC power plus the sum of the
active powers due to each harmonic frequency, which means
that the electrical magnitude with the narrower harmonic
distortion bandwidth determines the final bandwidth of the
power signal.
Revista elektron, Vol. 7, No. 1, pp. 19-27 (2023)
http://elektron.fi.uba.ar